null

RKI HS-04 H2S Single Gas Personal Gas Detector, 0-200 ppm

(No reviews yet) Write a Review
$195.00
$94.00
(You save 52% )
SKU:
RKI-73-0063
Availability:
FREE SHIPPING | Ships in 1-2 business days
  • Product Description

    RKI Instruments HS-04 Series H2S Single Gas Monitor is is a single-gas toxic personal monitor designed to operate for many years with field replaceable sensors and filters. With a protective rubber boot and an IP-67 rating, the 04 Series with a 3 year warranty is built to operate for a long time.

    Applications: Hazardous material, Mining Refineries / petrochemical, Water / wastewater treatment, Pharmaceuticals, Utilities, Chemical plants, Fire Services

    Key Features:

    • Personal monitoring for H2S
    • Operates 3,000 hours on replaceable alkaline batteries
    • Digital LCD readout
    • Easy "Turn On" and "Turn Off" by pressing down on the Power Mode button.
    • Pocket size 2.1"(W) x 2.6"(H) x 0.9"(D)
    • Light weight 2.8 ounces
    • Impact, dust, and water resistant (IP-67)
    • Replaceable rubber boot
    • Alligator clip with 360° rotation
    • Data log 10 to 600 hours with adjustable intervals
    • Audible / visual / vibration alarms
    • Automatic backlight
    • Low battery alarm
    • Intrinsically safe, ATEX and C / US approvals
    • Field serviceable
    • 3 year warranty on material and workmanship including sensors

    Specifications:

     

    • Gas Detected: Hydrogen Sulfide
    • Detection principle: Electrochemical
    • Detection range: 0 ~ 2000.00 ppm (1 ppm)
    • Alarm set points: 1st: 5 ppm, 2nd: 30 ppm, 3rd: 100 ppm, TWA: 1.0 ppm, STEL: 5.0 ppm
    • Alarm types: Gas alarms: 1st, 2nd and 3rd (user adjustable), OVER alarm, STEL and TWA, Trouble alarms: Sensor connection, low battery, circuit failure, and calibration range error
    • Gas sampling: Diffusion
    • Response time: T90 in 10 seconds
    • Accuracy: 0 - 100 ppm: ± 5% of reading or ± 2 ppm H2S (whichever is greater), 101 - 200 ppm: ± 20% of reading
    • Data logging: 10 second intervals = 5 hours, 5 minute intervals = 150 hours, 10 minute intervals = 300 hours Instrument records alarms, alarm trends, gas readings, calibration & bump test history, instrument information and settings.
    • Continuous operation: Alkaline dry cell: 3,000 hours (at 25°C, no backlight)
    • Safety rating: ATEX/IECEx approved. QPS classified, “C / US”, as intrinsically safe, Class I, Division 1, Groups A, B, C, D
    • Enclosure: Dust and water resistant with IP-67 rating
    • Display: Digital LCD readout, gas name & concentration, auto backlighting during alarm, operation icon, battery icon